CAL 03 – Triggering
The purpose of this section is to set the triggering levels for the 4 channels of the scope.
I didn’t capture photos of this, as it went by so fast, I forget to get snapshots. But all that showed was the normal display of Channel #, Input Voltage and Step number. There were no waveforms displayed.
The scope automatically runs steps 201 – 214 when you start the procedure.
Then in steps d – g, you are inputting a 500mv square wave to channel 1 and stepping through steps 215 and 216.
Steps h – k uses a 500mv square wave to Channel 2 to step through step 217.
Steps l – m uses a 500mv square wave to Channel 3 for step 218.
Steps n – o uses a 2 volt square wave to Channel 3 for step 219.
Steps p – q uses a 500mv square wave to Channel 4 for step 220.
Steps r – s uses a 2 volt square wave to Channel 4 for step 221.
It was all very straight forward…